NimOS Installation Guide

NimOS Installation Guide

NimOS Auto-Update


NimOS Software manual Installation

To get a NimOS installer, please raise a ticket to our Support team. 
  1. Once the installer has been downloaded, navigate to the download folder -- "C:\Users\ONI\Downloads" when accessed on the Nanoimager laptop. Double Click on the NimOS executable file (.exe). A new window will appear with the name Windows protected your PC. Click on More Info to proceed.

  2. A new button will appear with some additional information. Click on Run Anyway to start the NimOS installer.

  3. In the NimOS installer window, click on Next until it prompts the user if they would like to create a shortcut. Tick the checkbox to proceed with the installation. It is highly recommended to specify the patch number in the folder name. For example, "C:\Program Files\OxfordNanoimaging 1.19.1".
  4. During the installation process, the NimOS installer will install the required libraries and dependencies through the Windows command prompt. Once the installation is finished, you can start the NimOS software (Nanoimager.exe) from the desktop if Create A Desktop Shortcut is checked. Otherwise, you can start the NimOS software from the installation root folder.

NimOS Firmware Update

  1. In NimOS, Navigate to the Acquire tab and click on Connect To Microscope to connect to the microscope.
  2. In the process of connecting to the microscope, the system will check that the latest firmware has been installed. If not, the user will be prompted to install the latest version.
  3. If No is clicked, nothing will happen and NimOS will not connect to the microscope until the latest firmware is installed.
    If Yes is clicked, NimOS will install the latest firmware and a new window will appear to show the progress.
  4. Once NimOS has finished updating the firmware, click on Connect To Microscope again to reconnect to the microscope. If the installation of NimOS and the firmware was successful, no error message will appear and NimOS will connect to the microscope.

Python 3.6.6 Installation (if required)

The current NimOS software requires Python 3.6.6 to install the firmware. To check the Python version of your default environment, type python --version in the Windows command prompt and press Enter to print out the information:

If the Python version is 3.6.6 or later, skip to the NimOS Firmware Update.
If the Python version is not 3.6.6 or later, install Python 3.6.6:
  1. Download the required Python release from here (e.g. Python 3.6.6) and run the downloaded installer.
  2. In the Python installer window, check Add Python 3.6 to Path and click on Install Now to start the installation process.
  3. Once the installation is finished, confirm the Python version is correct in the command prompt and proceed with the firmware update. If the Python version is still not correct, check the System Environment Variables. Ensure that the root folder and the "Script" folder of Python 3.6.6 are at the top of the user Path list:


If at any point you are having issues with the installation process, please do not hesitate to contact the CX team.


    • Related Articles

    • NimOS Analysis

      NimOS provides functions for data analysis: Analysis: Tools Analysis: Viewing Options Analysis: Filter Analysis: Localizations Analysis: Drift Correction
    • NimOS Overview

      NimOS is the software used to control the microscope. Basic controls and terms are highlighted below. For video instructions on how to use NimOS to take an acquisition, check here. For NimOS tabs, check here. For laser control, check here. For stage ...
    • NimOS Tabs

      NimOS Tabs: Files: Load Data: open an acquisition, which will then appear in the Analyze interface. Instrument: Instrument Config: show general information about the microscope. Manual Connect: manually select an instrument file to connect to the ...
    • Camera calibration with NimOS 19.7 (and above)

      The procedure takes around 10 min INSTRUCTIONS: 1- Log in the Nanoimager laptop with an administrator account 2- If not already done, Download CODI Acquisition from the following link (1gb) here 3- Install CODI Acquisition and accept every prompt ...
    • Quick Guide: How to acquire dSTORM acquisition

      dSTORM image acquisition Start the Nanoimager and let the temperature stabilize to avoid drifts during acquisition. The stable microscope temperature is 6-7 C higher than room temperature(RT). To reach this stage quickly enter the desired (RT+ 6 C) ...